Intra-Aortic Balloon Pump (IABP) efficacy is critically affected by the inflation/deflation
timing. Balloon deflation may cause a sucking effect, and a steal phenomenon on carotid
flow. Delaying IABP deflation reduces the degree of this flow reversal, but at the
same time exposes patients to the risk of increased proto-systolic afterload with
detrimental effects on the LV. The purpose of this study was to investigate the effects
of a delayed IABP deflation timing on cerebral blood flow and LV hemodynamics, by
means of simultaneous carotid artery ultrasonography, trans-thoracic echocardiography
and central aortic pressure analysis. Delaying IABP deflation trigger to the beginning
of QRS effectively increased the cerebral blood flow by 20%, mostly by reducing the
reverse component flow caused by the diastolic balloon deflation. Extending the deflation
to the early systole was safe and favourably impacted on cardiac mechanics, increasing
CO by 15% without prolonging LV isovolumetric contraction and ejection phases.
KEYWORDS
Abbreviations:
ADP (augmented diastolic pressure), ASP (assisted systolic pressure), CO (cardiac output), EDP (end-diastolic pressure), IABP (intra-aortic balloon pump), LV (left ventricle), VTI (velocity-time integral)To read this article in full you will need to make a payment
Purchase one-time access:
Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online accessOne-time access price info
- For academic or personal research use, select 'Academic and Personal'
- For corporate R&D use, select 'Corporate R&D Professionals'
Subscribe:
Subscribe to The Journal of Heart and Lung TransplantationAlready a print subscriber? Claim online access
Already an online subscriber? Sign in
Register: Create an account
Institutional Access: Sign in to ScienceDirect
References
- Intra-Aortic Balloon pumping in acute decompensated heart failure with hypoperfusion: from pathophysiology to clinical practice.Circ Heart Fail. 2021; 14e008527
- Deflation timing influences Intra-Aortic Balloon Pump-Mediated carotid blood flow reversal: a case report.Circ Heart Fail. 2016; 9e003474
- Hemodynamic effects of new intra-aortic balloon counterpulsation timing methods in patients: a multicenter evaluation.Am Heart J. 1999; 137: 1129-1136
- Beat-to-beat effects of intraaortic balloon pump timing on left ventricular performance in patients with low ejection fraction.Ann Thorac Surg. 2005; 79: 872-880
- Intra-aortic balloon pump (IABP) counterpulsation improves cerebral perfusion in patients with decreased left ventricular function.Perfusion. 2014; 29: 511-516
Article info
Publication history
Published online: December 11, 2022
Publication stage
In Press Journal Pre-ProofIdentification
Copyright
© 2022 International Society for Heart and Lung Transplantation. All rights reserved.