Introduction and background
Heart failure (HF) remains a major cause of morbidity and mortality. Heart transplantation (HT) is a life-saving procedure for patients with advanced heart failure, but availability is limited by donor organ shortage.
1- Lund L.H.
- Edwards L.B.
- Kucheryavaya A.Y.
- et al.
The Registry of the International Society for Heart and Lung Transplantation: thirty-second official adult heart transplantation report—2015; Focus theme: Early graft failure.
Durable mechanical circulatory support (MCS) is a major advance in HF management and can provide hemodynamic support until myocardial recovery or HT, or as permanent therapy (destination therapy, DT) for those ineligible for HT.
2- Kirklin J.K.
- Naftel D.C.
- Pagani F.D.
- et al.
Seventh INTERMACS annual report: 15,000 patients and counting.
MCS has continued to enhance patient survival and quality of life through continued improvements in device design and durability.
2- Kirklin J.K.
- Naftel D.C.
- Pagani F.D.
- et al.
Seventh INTERMACS annual report: 15,000 patients and counting.
However, MCS infection remains as a major complication and a significant cause of morbidity and mortality in MCS recipients.
2- Kirklin J.K.
- Naftel D.C.
- Pagani F.D.
- et al.
Seventh INTERMACS annual report: 15,000 patients and counting.
Furthermore, there is considerable practice variation regarding patient selection for MCS, implant technique, infection prevention strategies anti-microbial prophylaxis, drive-line care and management of infection.
In 2015, the International Society for Heart and Lung Transplantation (ISHLT) recognized the importance of a consensus document on strategies for both prevention and management of MCS infections. The multidisciplinary consensus document is the result of a collaboration between the ISHLT and the International Consortium of Circulatory Assist Clinicians (ICCAC). The effort involved a number of ISHLT Councils, including: Infectious Disease; MCS; Nursing; Health Sciences and Allied Health (NHSAH); Heart Failure and Transplantation; Pharmacy; and Pediatric Thoracic Transplantation and Heart Failure. The panel members addressed the epidemiology and microbiology of MCS infections, pre-operative evaluation of the MCS candidate, peri-operative surgical and anti-microbial prophylaxis management, post-operative drive-line management and treatment of MCS infections.
The purpose of this document is to provide expert consensus-derived recommendations and, whenever possible, evidence--based recommendations for the prevention and management of infection in MCS recipients. Extensive literature review and the results of the ISHLT-supported international survey for infection control, prevention and management practices for MCS
3- Kusne S.
- Danziger-Isakov L.
- Mooney M.
- et al.
Infection control and prevention practices for mechanical circulatory support: an international survey.
were utilized to develop the consensus-derived recommendations presented here. The survey of infection control practices in ventricular assist devices (VADs) included detailed questions regarding infection control practices in VAD programs and was completed by 137 centers worldwide.
3- Kusne S.
- Danziger-Isakov L.
- Mooney M.
- et al.
Infection control and prevention practices for mechanical circulatory support: an international survey.
Two additional surveys of VAD infection control practices were also conducted (Kaan A et al, unpublished data, 2017).
4- Kaan A.
- Christensen D.
- Mackay M.
The real world: self-reported dressing practices in LVAD outpatients.
Proposed recommendations at the end of each section are ungraded due to the predominance of expert opinion.
The creation of the consensus document required multiple steps, but, briefly, the following was accomplished. The ISHLT Councils were contacted and names of experts interested in the project were suggested. After selecting a chair and co-chairs, the structure of the consensus document was established, and the writing teams were then chosen in collaboration with ISHLT councils. Face-to-face meetings and periodic conference calls were arranged and the writing process was then initiated. During this process, unclear areas and disagreements were identified. Because of lack of comparative trials, some of the statements in the consensus document required a decision based on majority vote. Several edits were required and, once the consensus document was created and reviewed by the authors, it was sent to external reviewers and revisions were incorporated.
In this document, MCS includes durable left ventricular assist devices (LVADs), right ventricular assist devices (RVADs), biventricular assist devices (BiVADs) and total artificial hearts (TAHs).
2- Kirklin J.K.
- Naftel D.C.
- Pagani F.D.
- et al.
Seventh INTERMACS annual report: 15,000 patients and counting.
These terms are used interchangeably throughout the document unless otherwise specified. Although the published literature pertains mostly to LVADs as the most widely used device type, the principles for infection prevention, diagnosis and management apply to all implantable and paracorporeal durable MCS devices with percutaneous drive-lines or cannulas, including pulsating and continuous-flow devices. It should be noted that there may be differences in infection rates and outcome between these devices, but the current literature does not provide any distinctions in infection prevention.
ISHLT definitions of infection
Standardized international definitions of MCS infections were developed in 2011 by an ISHLT multidisciplinary working group. Three categories of infection were defined for VAD patients: VAD-specific infections; VAD-related infections; and non-VAD infections.
5- Hannan M.M.
- Husain S.
- Mattner F.
- et al.
Working formulation for the standardization of definitions of infections in patients using ventricular assist devices.
These definitions were developed to allow for consistent data collection for each specific infection diagnosis and facilitate research in the field. A VAD-specific infection may involve any aspect of the device (pump, cannula, pocket or drive-line) and may cause sepsis. Infection may be introduced intra-operatively in the pump, cannula or pocket; may enter via the drive-line exit site; or may occur as a result of a bloodstream infection (BSI) from another focus of infection.
5- Hannan M.M.
- Husain S.
- Mattner F.
- et al.
Working formulation for the standardization of definitions of infections in patients using ventricular assist devices.
, 6- Gordon R.J.
- Quagliarello B.
- Lowy F.D.
Ventricular assist device-related infections.
, 7- Califano S.
- Pagani F.D.
- Malani P.N.
Left ventricular assist device-associated infections.
The VAD-specific infection definitions are each sub-divided into “proven, probable or possible” infection and include microbiologic, radiographic and clinical criteria for a “proven” diagnosis. The pump, cannula or pocket “proven” diagnosis is defined by intra-operative samples or needle aspiration of fluid from the MCS pocket. “Probable” or “possible” diagnoses are defined only by appropriate clinical assessment. The VAD-specific infections can be further divided into superficial drive-line or deep infections. A superficial drive-line infection involves soft tissue outside the fascia and muscle layers, whereas deep infection requires infection beyond these structures.
5- Hannan M.M.
- Husain S.
- Mattner F.
- et al.
Working formulation for the standardization of definitions of infections in patients using ventricular assist devices.
VAD-related infections are those that occur in patients without MCS but may be more common with presence of the device, such as endocarditis and mediastinitis.
5- Hannan M.M.
- Husain S.
- Mattner F.
- et al.
Working formulation for the standardization of definitions of infections in patients using ventricular assist devices.
Non-VAD infections are unrelated to the presence of the device (e.g., urinary tract infection) but are included as a category to provide a comprehensive description of infection in this population.
5- Hannan M.M.
- Husain S.
- Mattner F.
- et al.
Working formulation for the standardization of definitions of infections in patients using ventricular assist devices.
Standard surveillence definitions for these infections have been applied.
Epidemiology of infection in MCS
There are currently 3 large, multicenter MCS registries: (1) the Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS), a North American registry
; (2) the European Registry for Patients with Mechanical Circulatory Support (EUROMACS)
10European Registry for Patients with Mechanical Circulatory Support. www.euromacs.org/. 2017. Accessed January 10, 2017.
; and (3) the ISHLT Registry for Mechanically Assisted Circulatory Support (IMACS).
These registries collect and report data on major infections, and are partly linked to each other (e.g., IMACS includes INTERMACS and EUROMACS data). Infection occurs in up to 60% of MCS recipients with the increased rate related to duration of device placement.
6- Gordon R.J.
- Quagliarello B.
- Lowy F.D.
Ventricular assist device-related infections.
, 7- Califano S.
- Pagani F.D.
- Malani P.N.
Left ventricular assist device-associated infections.
, 12- de By T.M.
- Mohacsi P.
- Gummert J.
- et al.
The European Registry for Patients with Mechanical Circulatory Support (EUROMACS): first annual report.
In many reports it is difficult to determine the true incidence of MCS-associated infections because standardized definitions were only recently employed, and earlier databases lack detailed infectious disease--related variables.
Since the mid-2000s there have been 2 major changes in LVAD practice. First, smaller continuous-flow devices have replaced larger pulsatile flow pumps and currently represent >95% of LVAD implants. The smaller pocket and drive-line dimensions of continuous-flow devices have been associated with decreased LVAD infection rates.
2- Kirklin J.K.
- Naftel D.C.
- Pagani F.D.
- et al.
Seventh INTERMACS annual report: 15,000 patients and counting.
, 13- Kirklin J.K.
- Naftel D.C.
- Kormos R.L.
- et al.
Fifth INTERMACS annual report: risk factor analysis from more than 6,000 mechanical circulatory support patients.
, 14- Topkara V.K.
- Kondareddy S.
- Malik F.
- et al.
Infectious complications in patients with left ventricular assist device: etiology and outcomes in the continuous-flow era.
Second, LVAD use for DT has increased over time.
2- Kirklin J.K.
- Naftel D.C.
- Pagani F.D.
- et al.
Seventh INTERMACS annual report: 15,000 patients and counting.
, 13- Kirklin J.K.
- Naftel D.C.
- Kormos R.L.
- et al.
Fifth INTERMACS annual report: risk factor analysis from more than 6,000 mechanical circulatory support patients.
, 15- Kirklin J.K.
- Naftel D.C.
- Pagani F.D.
- et al.
Sixth INTERMACS annual report: a 10,000-patient database.
In the most recent INTERMACS data, DT accounted for 45.7% of MCS implants and 100% of these were continuous-flow devices.
2- Kirklin J.K.
- Naftel D.C.
- Pagani F.D.
- et al.
Seventh INTERMACS annual report: 15,000 patients and counting.
The fifth INTERMACS report included data on 6,561 patients and demonstrated that infection was considerably less common with continuous-flow devices. It also showed that pump interior infections and pocket infections were uncommon events, but the risk of drive-line infection persisted as long as patients were monitored.
13- Kirklin J.K.
- Naftel D.C.
- Kormos R.L.
- et al.
Fifth INTERMACS annual report: risk factor analysis from more than 6,000 mechanical circulatory support patients.
The seventh INTERMACS report included data on >15,000 patients who received a primary implant from June 2006 to December 2014.
2- Kirklin J.K.
- Naftel D.C.
- Pagani F.D.
- et al.
Seventh INTERMACS annual report: 15,000 patients and counting.
Infection was the fourth most common cause of death within 1 year after implant (228 of 9,781, or 8.8%), following neurologic complications (15.6%), multisystem organ failure (MSOF) (15.6%) and withdrawal of support (10.4%). The infection incidence may be higher than reported as it is not clear whether infection was an important factor in patients with MSOF or those with withdrawn support. Although risk of death from neurologic complications remained stable in the second year post-implantation, the risk of infection-associated death gradually increased over the monitored time period and became a major cause of late mortality (together with neurologic and MSOF).
2- Kirklin J.K.
- Naftel D.C.
- Pagani F.D.
- et al.
Seventh INTERMACS annual report: 15,000 patients and counting.
The first EUROMACS report included data on 741 patients. Continuous-flow devices predominated, as with INTERMACS, and infection was the most common major adverse event in this cohort. Of the 433 major adverse events, 153 (35.3%) were due to major infections, which represented the second most common cause of death. Of the 293 deaths, 69 (23.5%) were caused by infection or sepsis.
12- de By T.M.
- Mohacsi P.
- Gummert J.
- et al.
The European Registry for Patients with Mechanical Circulatory Support (EUROMACS): first annual report.
There is a subset of advanced HF patients who require BiVAD or TAH as bridge to transplant (BTT). In addition to BTT, the United States Food and Drug Administration has approved 1 biventricular device (Syncardia TAH) for investigational use in DT.
16SynCardia Systems I. Press Release: FDA approves the SynCardia Total Artificial Heart for Destination Therapy Study. http://www.syncardia.com. Accessed on July 1, 2016.
However, there are other BiVADs under investigation and used elsewhere. One report described 101 TAH patients with a median support of 53 days (range 1 to 441 days).
17- Copeland J.G.
- Copeland H.
- Gustafson M.
- et al.
Experience with more than 100 total artificial heart implants.
In that study there was a 63% infection rate, with 50% of cases occurring in the first 30 days post-implant. Late complications were evaluated in 47 TAH patients beyond 1 year from 10 centers, with 53% and 27% of these patients having systemic and drive-line infection, respectively.
18- Torregrossa G.
- Morshuis M.
- Varghese R.
- et al.
Results with SynCardia total artificial heart beyond 1 year.
Infection and hemorrhagic events (intracranial and gastrointestinal) were the major causes of death. Mediastinitis developed in up to 4% of LVAD and TAH recipients due to direct spread from drive-line infection.
17- Copeland J.G.
- Copeland H.
- Gustafson M.
- et al.
Experience with more than 100 total artificial heart implants.
, 18- Torregrossa G.
- Morshuis M.
- Varghese R.
- et al.
Results with SynCardia total artificial heart beyond 1 year.
, 19- Copeland J.G.
- Smith R.G.
- Arabia F.A.
- et al.
Total artificial heart bridge to transplantation: a 9-year experience with 62 patients.
, 20- Arabia F.A.
- Copeland J.G.
- Smith R.G.
- et al.
Infections with the CardioWest total artificial heart.
, 21- Taimur S.
- Sullivan T.
- Rana M.
Successful heart transplantation in patients with total artificial heart infection.
Microbiology
Despite the emergence of continuous-flow devices and increased DT use, which has changed patient profiles and increased the duration of MCS support, pathogens associated with MCS infection have not changed over time.
22- Weyand M.
- Hermann M.
- Kondruweit M.
- et al.
Clinical impact of infections in left ventricular assist device recipients: the importance of site and organism.
, 23- Gordon S.M.
- Schmitt S.K.
- Jacobs M.
- et al.
Nosocomial bloodstream infections in patients with implantable left ventricular assist devices.
, 24- Nienaber J.J.
- Kusne S.
- Riaz T.
- et al.
Clinical manifestations and management of left ventricular assist device-associated infections.
, 25- Gordon R.J.
- Weinberg A.D.
- Pagani F.D.
- et al.
Prospective, multicenter study of ventricular assist device infections.
, 26- Schaffer J.M.
- Allen J.G.
- Weiss E.S.
- et al.
Infectious complications after pulsatile-flow and continuous-flow left ventricular assist device implantation.
Bacteria are the dominant etiology of early and late infections, and fungi are less prevalent (<10%).
23- Gordon S.M.
- Schmitt S.K.
- Jacobs M.
- et al.
Nosocomial bloodstream infections in patients with implantable left ventricular assist devices.
, 24- Nienaber J.J.
- Kusne S.
- Riaz T.
- et al.
Clinical manifestations and management of left ventricular assist device-associated infections.
, 25- Gordon R.J.
- Weinberg A.D.
- Pagani F.D.
- et al.
Prospective, multicenter study of ventricular assist device infections.
, 26- Schaffer J.M.
- Allen J.G.
- Weiss E.S.
- et al.
Infectious complications after pulsatile-flow and continuous-flow left ventricular assist device implantation.
, 27- Sharma V.
- Deo S.V.
- Stulak J.M.
- et al.
Driveline infections in left ventricular assist devices: implications for destination therapy.
, 28- Zierer A.
- Melby S.J.
- Voeller R.K.
- et al.
Late-onset driveline infections: the Achilles׳ heel of prolonged left ventricular assist device support.
, 29- Nienaber J.
- Wilhelm M.P.
- Sohail M.R.
Current concepts in the diagnosis and management of left ventricular assist device infections.
, 30- Toba F.A.
- Akashi H.
- Arrecubieta C.
- et al.
Role of biofilm in Staphylococcus aureus and Staphylococcus epidermidis ventricular assist device driveline infections.
, 31- Nurozler F.
- Argenziano M.
- Oz M.C.
- et al.
Fungal left ventricular assist device endocarditis.
, 32- Shoham S.
- Shaffer R.
- Sweet L.
- et al.
Candidemia in patients with ventricular assist devices.
, 33- Bagdasarian N.G.
- Malani A.N.
- Pagani F.D.
- et al.
Fungemia associated with left ventricular assist device support.
, 34- Maly J.
- Szarszoi O.
- Netuka I.
- et al.
Fungal infections associated with long-term mechanical circulatory support-diagnosis and management.
, 35Epidemiology and control of nosocomial infections in adult intensive care units.
The most common pathogens are Gram-positive bacteria that colonize skin and adhere to implanted material and create biofilm,
30- Toba F.A.
- Akashi H.
- Arrecubieta C.
- et al.
Role of biofilm in Staphylococcus aureus and Staphylococcus epidermidis ventricular assist device driveline infections.
particularly
Staphylococcus aureus and
Staphylococcus epidermidis, causing >50% of MCS infections.
Enterococcus species are the third most common Gram-positive organism (2%). The most common Gram-negative bacteria is
Pseudomonas aeruginosa (22% to 28%), followed by Enterobacteriacea such as
Klebsiella (2% to 4%) and
Enterobacter (2%).
22- Weyand M.
- Hermann M.
- Kondruweit M.
- et al.
Clinical impact of infections in left ventricular assist device recipients: the importance of site and organism.
, 23- Gordon S.M.
- Schmitt S.K.
- Jacobs M.
- et al.
Nosocomial bloodstream infections in patients with implantable left ventricular assist devices.
, 24- Nienaber J.J.
- Kusne S.
- Riaz T.
- et al.
Clinical manifestations and management of left ventricular assist device-associated infections.
, 25- Gordon R.J.
- Weinberg A.D.
- Pagani F.D.
- et al.
Prospective, multicenter study of ventricular assist device infections.
, 26- Schaffer J.M.
- Allen J.G.
- Weiss E.S.
- et al.
Infectious complications after pulsatile-flow and continuous-flow left ventricular assist device implantation.
, 27- Sharma V.
- Deo S.V.
- Stulak J.M.
- et al.
Driveline infections in left ventricular assist devices: implications for destination therapy.
, 28- Zierer A.
- Melby S.J.
- Voeller R.K.
- et al.
Late-onset driveline infections: the Achilles׳ heel of prolonged left ventricular assist device support.
It is important for clinicians to be familiar with local geographic and institutional microbiology and anti-microbial resistance patterns. The National Healthcare Safety Network’s (associated with the Centers for Disease Control and Prevention) surveillance of antibiotic-resistant pathogens in surgical site infections (SSIs) showed a decrease of 9.0% in methicillin-resistant
S aureus (MRSA)-caused SSIs between 2007 to 2008 and 2009 to 2010. However, there was a substantial increase in other resistant organisms, including vancomycin-resistant
Enterococcus (VRE, by 34.7%), and multidrug-resistant Gram-negatives, including
Escherichia coli (by 41.8%),
Pseudomonas (by 8.4%) and
Enterobacter (by 10.6%).
36- Sievert D.M.
- Ricks P.
- Edwards J.R.
- et al.
Antimicrobial-resistant pathogens associated with healthcare-associated infections: summary of data reported to the National Healthcare Safety Network at the Centers for Disease Control and Prevention, 2009-2010.
Although uncommon, fungal infections may also occur and are extremely difficult to eradicate.
31- Nurozler F.
- Argenziano M.
- Oz M.C.
- et al.
Fungal left ventricular assist device endocarditis.
, 32- Shoham S.
- Shaffer R.
- Sweet L.
- et al.
Candidemia in patients with ventricular assist devices.
, 33- Bagdasarian N.G.
- Malani A.N.
- Pagani F.D.
- et al.
Fungemia associated with left ventricular assist device support.
, 34- Maly J.
- Szarszoi O.
- Netuka I.
- et al.
Fungal infections associated with long-term mechanical circulatory support-diagnosis and management.
Candidemia has been reported in 1.3% to 9.7% of MCS recipients, with an overall mortality of 15% to 25%.
33- Bagdasarian N.G.
- Malani A.N.
- Pagani F.D.
- et al.
Fungemia associated with left ventricular assist device support.
, 34- Maly J.
- Szarszoi O.
- Netuka I.
- et al.
Fungal infections associated with long-term mechanical circulatory support-diagnosis and management.
Candida albicans is the most common fungal pathogen (70%) in MCS infection, followed by
Candida glabrata (10%).
Aspergillus species have also been identified as a rare cause of life-threatening MCS infections, but there are limited data regarding incidence. In most reported cases,
Aspergillus device infection was detected post-mortem.
34- Maly J.
- Szarszoi O.
- Netuka I.
- et al.
Fungal infections associated with long-term mechanical circulatory support-diagnosis and management.
Risk factors for infection
Clinicians evaluating patients for MCS should be aware of the risk factors for infection in this population. The majority of studies evaluating risk factors were retrospective cohort studies with a predominance of pulsatile MCS devices. Risk factors for MCS-related infections include older age,
26- Schaffer J.M.
- Allen J.G.
- Weiss E.S.
- et al.
Infectious complications after pulsatile-flow and continuous-flow left ventricular assist device implantation.
diabetes,
37- Bucerius J.
- Gummert J.F.
- Walther T.
- et al.
Impact of diabetes mellitus on cardiac surgery outcome.
, 38- Topkara V.K.
- Dang N.C.
- Martens T.P.
- et al.
Effect of diabetes on short- and long-term outcomes after left ventricular assist device implantation.
, 39- John R.
- Aaronson K.D.
- Pae W.E.
- et al.
Drive-line infections and sepsis in patients receiving the HVAD system as a left ventricular assist device.
renal failure,
25- Gordon R.J.
- Weinberg A.D.
- Pagani F.D.
- et al.
Prospective, multicenter study of ventricular assist device infections.
greater severity of heart failure,
26- Schaffer J.M.
- Allen J.G.
- Weiss E.S.
- et al.
Infectious complications after pulsatile-flow and continuous-flow left ventricular assist device implantation.
malnutrition,
40- Holdy K.
- Dembitsky W.
- Eaton L.L.
- et al.
Nutrition assessment and management of left ventricular assist device patients.
, 41- Imamura T.
- Kinugawa K.
- Nitta D.
- et al.
Readmission due to driveline infection can be predicted by new score by using serum albumin and body mass index during long-term left ventricular assist device support.
T-cell dysfunction associated with the device,
42- Kimball P.M.
- Flattery M.
- McDougan F.
- et al.
Cellular immunity impaired among patients on left ventricular assist device for 6 months.
hypogammaglobulinemia,
43- Yamani M.H.
- Chuang H.H.
- Ozduran V.
- et al.
The impact of hypogammaglobulinemia on infection outcome in patients undergoing ventricular assist device implantation.
intravascular lines and organisms capable of biofilm formation,
44Infection in ventricular assist devices: the role of biofilm.
, 45- Gristina A.G.
- Dobbins J.J.
- Giammara B.
- et al.
Biomaterial-centered sepsis and the total artificial heart. Microbial adhesion vs tissue integration.
, 46Biofilm infections, their resilience to therapy and innovative treatment strategies.
obesity,
39- John R.
- Aaronson K.D.
- Pae W.E.
- et al.
Drive-line infections and sepsis in patients receiving the HVAD system as a left ventricular assist device.
, 42- Kimball P.M.
- Flattery M.
- McDougan F.
- et al.
Cellular immunity impaired among patients on left ventricular assist device for 6 months.
delayed sternal closure,
47- Stulak J.M.
- Romans T.
- Cowger J.
- et al.
Delayed sternal closure does not increase late infection risk in patients undergoing left ventricular assist device implantation.
longer intensive care unit (ICU) stay
48- Poston R.S.
- Husain S.
- Sorce D.
- et al.
LVAD bloodstream infections: therapeutic rationale for transplantation after LVAD infection.
and prolonged duration of MCS support.
28- Zierer A.
- Melby S.J.
- Voeller R.K.
- et al.
Late-onset driveline infections: the Achilles׳ heel of prolonged left ventricular assist device support.
, 48- Poston R.S.
- Husain S.
- Sorce D.
- et al.
LVAD bloodstream infections: therapeutic rationale for transplantation after LVAD infection.
Infection risk factor analysis specifically for continuous-flow devices recipients is limited. In an analysis of 332 BTT HeartWare recipients, risk factors for drive-line infection and sepsis included larger body mass index (BMI) and diabetes.
39- John R.
- Aaronson K.D.
- Pae W.E.
- et al.
Drive-line infections and sepsis in patients receiving the HVAD system as a left ventricular assist device.
A prospective study that included continuous-flow VAD devices identified elevated serum creatinine and history of depression as independent risk factors for infection.
25- Gordon R.J.
- Weinberg A.D.
- Pagani F.D.
- et al.
Prospective, multicenter study of ventricular assist device infections.
There was no difference in risk factors for infection between pulsatile- and continuous-flow devices, or between BTT and DT. Neither length of time spent in ICU nor total hospital length of stay was associated with increased infection in that study.
25- Gordon R.J.
- Weinberg A.D.
- Pagani F.D.
- et al.
Prospective, multicenter study of ventricular assist device infections.
Trauma to the drive-line has been identified as a major risk factor for development of drive-line infection in both pulsatile- and continuous-flow devices. Accidental pulling of the drive-line and disruption of the seal between drive-line and surrounding skin are commonly reported causes of infection.
28- Zierer A.
- Melby S.J.
- Voeller R.K.
- et al.
Late-onset driveline infections: the Achilles׳ heel of prolonged left ventricular assist device support.
In contrast, 1 single-center, retrospective study of 194 HeartMate II recipients, showed no association between drive-line infection and the following factors: BMI; age; education level; insurance provider; and velour position at exit site.
49- Koval C.E.
- Thuita L.
- Moazami N.
- et al.
Evolution and impact of drive-line infection in a large cohort of continuous-flow ventricular assist device recipients.
Another study,
50- Haglund N.A.
- Davis M.E.
- Tricarico N.M.
- et al.
Readmissions after continuous flow left ventricular assist device implantation: differences observed between two contemporary device types.
which followed BTT patients between 2009 and 2014, demonstrated that the number of re-admissions for infection was higher in HeartMate II compared with HeartWare HVAD recipients (both continuous-flow devices).
50- Haglund N.A.
- Davis M.E.
- Tricarico N.M.
- et al.
Readmissions after continuous flow left ventricular assist device implantation: differences observed between two contemporary device types.
It is expected that a larger device requiring pocket creation in HeartMate II would be prone to higher infection rate compared with the smaller device without a pocket creation in HeartWare HVAD. However, the ENDURANCE trial showed no difference in infection rate between the HeartMate II and the HVAD.
51- Rogers J.G.
- Pagani F.D.
- Tatooles A.J.
- et al.
Intrapericardial left ventricular assist device for advanced heart failure.
Clinical outcomes
In the current era of continuous-flow devices, the short- and long-term survival outcomes after HT in patients bridged with MCS are similar to those without pre-HT MCS.
52- Kamdar F.
- John R.
- Eckman P.
- et al.
Postcardiac transplant survival in the current era in patients receiving continuous-flow left ventricular assist devices.
, 53- Urban M.
- Pirk J.
- Dorazilova Z.
- et al.
How does successful bridging with ventricular assist device affect cardiac transplantation outcome?.
, 54- Hequet D.
- Kralidis G.
- Carrel T.
- et al.
Ventricular assist devices as bridge to heart transplantation: impact on post-transplant infections.
A “best evidence topic in cardiac surgery” report reviewed 428 articles studying the impact of BTT on survival after HT.
53- Urban M.
- Pirk J.
- Dorazilova Z.
- et al.
How does successful bridging with ventricular assist device affect cardiac transplantation outcome?.
In that study it was concluded that patients’ survival of BTT followed by HT was comparable to that seen in those with HT who did not have MCS beforehand.
53- Urban M.
- Pirk J.
- Dorazilova Z.
- et al.
How does successful bridging with ventricular assist device affect cardiac transplantation outcome?.
In patients who did not have HT, those who developed infection had poorer prognosis.
55- Goldstein D.J.
- Naftel D.
- Holman W.
- et al.
Continuous-flow devices and percutaneous site infections: clinical outcomes.
MCS infection is associated with an increased rate of re-hospitalization, morbidity and mortality if MCS is not followed by HT.
28- Zierer A.
- Melby S.J.
- Voeller R.K.
- et al.
Late-onset driveline infections: the Achilles׳ heel of prolonged left ventricular assist device support.
, 46Biofilm infections, their resilience to therapy and innovative treatment strategies.
, 48- Poston R.S.
- Husain S.
- Sorce D.
- et al.
LVAD bloodstream infections: therapeutic rationale for transplantation after LVAD infection.
, 49- Koval C.E.
- Thuita L.
- Moazami N.
- et al.
Evolution and impact of drive-line infection in a large cohort of continuous-flow ventricular assist device recipients.
, 56- Tong M.Z.
- Smedira N.G.
- Soltesz E.G.
- et al.
Outcomes of heart transplant after left ventricular assist device specific and related infection.
Superficial drive-line infection (DLI) may progress over time to become a deep tissue infection.
49- Koval C.E.
- Thuita L.
- Moazami N.
- et al.
Evolution and impact of drive-line infection in a large cohort of continuous-flow ventricular assist device recipients.
, 57- Schulman A.R.
- Martens T.P.
- Russo M.J.
- et al.
Effect of left ventricular assist device infection on post-transplant outcomes.
Patients with deep tissue infection were found to be more likely to develop bacteremia.
49- Koval C.E.
- Thuita L.
- Moazami N.
- et al.
Evolution and impact of drive-line infection in a large cohort of continuous-flow ventricular assist device recipients.
Sepsis is the most common cause of death associated with DLI. The organisms causing sepsis may be different from those causing DLI, and patients on suppressive therapy after DLI may develop secondary infection.
39- John R.
- Aaronson K.D.
- Pae W.E.
- et al.
Drive-line infections and sepsis in patients receiving the HVAD system as a left ventricular assist device.
, 49- Koval C.E.
- Thuita L.
- Moazami N.
- et al.
Evolution and impact of drive-line infection in a large cohort of continuous-flow ventricular assist device recipients.
, 55- Goldstein D.J.
- Naftel D.
- Holman W.
- et al.
Continuous-flow devices and percutaneous site infections: clinical outcomes.
In the INTERMACS registry, 22% of deaths among those who developed DLI occurred within 1 month of DLI diagnosis, suggesting a potential association.
2- Kirklin J.K.
- Naftel D.C.
- Pagani F.D.
- et al.
Seventh INTERMACS annual report: 15,000 patients and counting.
, 55- Goldstein D.J.
- Naftel D.
- Holman W.
- et al.
Continuous-flow devices and percutaneous site infections: clinical outcomes.
Infection is the fourth most common cause of death within 1 year, and the risk of infection was shown to rise steadily over a 4-year follow-up.
2- Kirklin J.K.
- Naftel D.C.
- Pagani F.D.
- et al.
Seventh INTERMACS annual report: 15,000 patients and counting.
Numerous studies have shown that patients with MCS infection can be transplanted successfully, with outcomes similar to those without MCS infection.
28- Zierer A.
- Melby S.J.
- Voeller R.K.
- et al.
Late-onset driveline infections: the Achilles׳ heel of prolonged left ventricular assist device support.
, 39- John R.
- Aaronson K.D.
- Pae W.E.
- et al.
Drive-line infections and sepsis in patients receiving the HVAD system as a left ventricular assist device.
, 46Biofilm infections, their resilience to therapy and innovative treatment strategies.
, 48- Poston R.S.
- Husain S.
- Sorce D.
- et al.
LVAD bloodstream infections: therapeutic rationale for transplantation after LVAD infection.
, 49- Koval C.E.
- Thuita L.
- Moazami N.
- et al.
Evolution and impact of drive-line infection in a large cohort of continuous-flow ventricular assist device recipients.
, 55- Goldstein D.J.
- Naftel D.
- Holman W.
- et al.
Continuous-flow devices and percutaneous site infections: clinical outcomes.
, 58- Toda K.
- Yonemoto Y.
- Fujita T.
- et al.
Risk analysis of bloodstream infection during long-term left ventricular assist device support.
, 59- Quader M.A.
- Wolfe L.G.
- Kasirajan V.
Heart transplantation outcomes in patients with continuous-flow left ventricular assist device-related complications.
A recent report has contradicted this assumption. A review of United Network for Organ Sharing (UNOS) data from 2006 to 2012 noted increased mortality in HT recipients after MCS-related infection compared with MCS without infection, showing survival rates of 85.6% vs 89.9% and 78.0% vs 82.7% at 1 and 3 years (
p = 0.01), respectively.
59- Quader M.A.
- Wolfe L.G.
- Kasirajan V.
Heart transplantation outcomes in patients with continuous-flow left ventricular assist device-related complications.
The INTERMACS registry followed 301 TAH recipients and noted no change over time in 1-year survival (59%) in the periods 2008 to 2011 and 2012 to 2014.
2- Kirklin J.K.
- Naftel D.C.
- Pagani F.D.
- et al.
Seventh INTERMACS annual report: 15,000 patients and counting.
In 1 single-center report, 7 infected TAH recipients underwent HT, 5 of whom were alive at a median time of 181 days and 2 deaths secondary to intra-operative bleed and cardiac arrest. The successful HT outcome in infected TAH patients was assumed to result from meticulous pre- and post-HT medical and surgical management.
19- Copeland J.G.
- Smith R.G.
- Arabia F.A.
- et al.
Total artificial heart bridge to transplantation: a 9-year experience with 62 patients.
Patient selection for HT
60- Mehra M.R.
- Canter C.E.
- Hannan M.M.
- et al.
The 2016 International Society for Heart Lung Transplantation listing criteria for heart transplantation: a 10-year update.
and post-HT management is considered essential to successful HT in patients with MCS infection.
19- Copeland J.G.
- Smith R.G.
- Arabia F.A.
- et al.
Total artificial heart bridge to transplantation: a 9-year experience with 62 patients.
, 60- Mehra M.R.
- Canter C.E.
- Hannan M.M.
- et al.
The 2016 International Society for Heart Lung Transplantation listing criteria for heart transplantation: a 10-year update.
Infectious disease evaluation of MCS candidates
General considerations
Formal infectious disease (ID) consultation may be considered based on individual circumstances, but it is not routinely required. However, all MCS candidates should be evaluated for active infection that would need to be addressed before surgery and for identification of modifiable risk factors for infection post-MCS. In BTT candidates, the heart transplant ID evaluation may be initiated at this time. Patient-specific factors should guide the extent and frequency of ID evaluation.
Infection should be excluded or appropriately treated before MCS implantation when clinically feasible. Evaluation of suspected infection in potential MCS recipients is not different from that in other patients and should be guided by clinical signs and symptoms. In patients with unexplained fever and/or leukocytosis, evaluation should include blood cultures, urinalysis, urine culture and chest X-ray, with additional imaging as needed until a diagnosis is established and the source has been treated and cleared.
ID consultation should be performed in all MCS candidates with suspected or proven infection. MCS candidates with bloodstream infections should be treated with targeted anti-microbial therapy. A consideration for delaying MCS implantation is recommended until the following goals are met: source control (e.g., incision and drainage of abscess, removal of infected catheter or tooth extraction for dental abscess); negative blood cultures; and resolution of illness and sepsis. MCS candidates with other infections (e.g., pneumonia, urinary tract infection) should be treated with appropriate anti-microbial therapy until resolution. There is insufficient evidence to define a minimum duration of anti-microbial therapy for an active infection before proceeding to MCS implantation.
Nosocomial infection and indwelling catheters
Nosocomial BSI is a major cause of morbidity and mortality after MCS placement.
61- Haglund N.A.
- Cox Z.L.
- Lee J.T.
- et al.
Are peripherally inserted central catheters associated with increased risk of adverse events in status 1B patients awaiting transplantation on continuous intravenous milrinone?.
Indwelling lines are a major source of nosocomial infection. By definition, long-term catheters are those indwelling for ≥14 days.
62- Mermel L.A.
- Allon M.
- Bouza E.
- et al.
Clinical practice guidelines for the diagnosis and management of intravascular catheter-related infection: 2009 update by the Infectious Diseases Society of America.
The infection risk associated with catheters varies based on host factors (age, immune status) and catheter type, location and duration.
63- Maki D.G.
- Kluger D.M.
- Crnich C.J.
The risk of bloodstream infection in adults with different intravascular devices: a systematic review of 200 published prospective studies.
Although all intravascular catheters are a potential source of BSI, temporary non-cuffed/non-tunneled central venous catheters (CVCs) are associated with a higher risk of BSI compared with peripherally inserted central catheters (PICCs) and cuffed/tunneled central catheters.
62- Mermel L.A.
- Allon M.
- Bouza E.
- et al.
Clinical practice guidelines for the diagnosis and management of intravascular catheter-related infection: 2009 update by the Infectious Diseases Society of America.
, 63- Maki D.G.
- Kluger D.M.
- Crnich C.J.
The risk of bloodstream infection in adults with different intravascular devices: a systematic review of 200 published prospective studies.
, 64- Chopra V.
- O׳Horo J.C.
- Rogers M.A.
- et al.
The risk of bloodstream infection associated with peripherally inserted central catheters compared with central venous catheters in adults: a systematic review and meta-analysis.
, 65- Miller S.E.
- Maragakis L.L.
Central line-associated bloodstream infection prevention.
, 66- Miller R.S.
- Norris P.R.
- Jenkins J.M.
- et al.
Systems initiatives reduce healthcare-associated infections: a study of 22,928 device days in a single trauma unit.
, 67- O׳Grady N.P.
- Alexander M.
- Burns L.A.
- et al.
Guidelines for the prevention of intravascular catheter-related infections.
CVCs placed in femoral locations carry a higher infection risk than jugular and subclavian locations.
67- O׳Grady N.P.
- Alexander M.
- Burns L.A.
- et al.
Guidelines for the prevention of intravascular catheter-related infections.
, 68- Maki D.G.
- Goldman D.A.
- Rhame F.S.
Infection control in intravenous therapy.
Most importantly, catheter duration has consistently been shown to be a risk factor for BSI. Although not studied specifically in the MCS population, taking a proactive approach to eliminate unnecessary catheters has been shown to reduce BSI in the ICU and post-surgical settings.
65- Miller S.E.
- Maragakis L.L.
Central line-associated bloodstream infection prevention.
, 66- Miller R.S.
- Norris P.R.
- Jenkins J.M.
- et al.
Systems initiatives reduce healthcare-associated infections: a study of 22,928 device days in a single trauma unit.
, 67- O׳Grady N.P.
- Alexander M.
- Burns L.A.
- et al.
Guidelines for the prevention of intravascular catheter-related infections.
, 69- Marsteller J.A.
- Sexton J.B.
- Hsu Y.J.
- et al.
A multicenter, phased, cluster-randomized controlled trial to reduce central line-associated bloodstream infections in intensive care units.
, 70- Huang E.Y.
- Chen C.
- Abdullah F.
- et al.
Strategies for the prevention of central venous catheter infections: an American Pediatric Surgical Association Outcomes and Clinical Trials Committee systematic review.
Catheter-associated urinary tract infection (CAUTI) is the most common nosocomial infection and is one that is preventable
71- Hooton T.M.
- Bradley S.F.
- Cardenas D.D.
- et al.
Diagnosis, prevention, and treatment of catheter-associated urinary tract infection in adults: 2009 International Clinical Practice Guidelines from the Infectious Diseases Society of America.
by limiting the number of days of catheterization. One study showed a 3-fold increased mortality risk in hospitalized patients who had required an indwelling urinary catheter.
72- Platt R.
- Polk B.F.
- Murdock B.
- et al.
Mortality associated with nosocomial urinary-tract infection.
In 2 other studies, CAUTI was found to be the second most common cause of nosocomial bacteremia
73Engineering out the risk for infection with urinary catheters.
and the most common cause of Gram-negative bacteremia.
74- Kreger B.E.
- Craven D.E.
- Carling P.C.
- et al.
Gram-negative bacteremia. III. Reassessment of etiology, epidemiology and ecology in 612 patients.
This is important because seeding of the device may occur in MCS recipients.
Strategies to reduce the risk of CAUTI include: placement of indwelling catheter only if indicated, not for routine incontinence; development of a list of legitimate indications together with staff education; requirement of a physician’s signature for each indwelling catheter placement; and consideration of a bladder scanner use to check if catheterization is necessary post-operatively.
71- Hooton T.M.
- Bradley S.F.
- Cardenas D.D.
- et al.
Diagnosis, prevention, and treatment of catheter-associated urinary tract infection in adults: 2009 International Clinical Practice Guidelines from the Infectious Diseases Society of America.
Colonization
Based on the ISHLT-supported survey, 94% of VAD centers perform nasal colonization surveillance culture for MRSA and 28% perform rectal VRE colonization surveillance in MCS candidates.
3- Kusne S.
- Danziger-Isakov L.
- Mooney M.
- et al.
Infection control and prevention practices for mechanical circulatory support: an international survey.
Ninety-two percent of centers use antiseptics to reduce colonizing bacteria before MCS implant surgery.
3- Kusne S.
- Danziger-Isakov L.
- Mooney M.
- et al.
Infection control and prevention practices for mechanical circulatory support: an international survey.
Mupirocin nasal decontamination has been shown to decrease
S aureus infections in surgery patients.
75- Kallen A.J.
- Wilson C.T.
- Larson R.J.
Perioperative intranasal mupirocin for the prevention of surgical-site infections: systematic review of the literature and meta-analysis.
, 76Decolonization therapy in infection control.
Although there have been no specific studies in MCS recipients, many centers have adapted this practice. Daily bathing with chlorhexidine using disposable washcloths has been shown to reduce the rates of hospital-acquired infection with multidrug-resistant organisms.
77- Climo M.W.
- Yokoe D.S.
- Warren D.K.
- et al.
Effect of daily chlorhexidine bathing on hospital-acquired infection.
Pre-operative chlorhexidine bathing has been shown to reduce skin bacterial counts but not post-operative wound infection rates after cardiac surgery.
78- Lazar H.L.
- Salm T.V.
- Engelman R.
- et al.
Prevention and management of sternal wound infections.
According to The Society for Healthcare Epidemiology of America/Infectious Diseases Society of America practice recommendations, there are not enough data to support routine pre-operative chlorhexidine bathing.
79- Anderson D.J.
- Podgorny K.
- Berrios-Torres S.I.
- et al.
Strategies to prevent surgical site infections in acute care hospitals: 2014 update.
The recent 2017 CDC guidelines for prevention of SSIs recommend showering or bathing with soap or an antiseptic agent at least the night before surgery.
80Berrios-Torres SI, Umscheid CA, Bratzler DW, et al. Centers for Disease Control and Prevention Guideline for the Prevention of Surgical Site Infection [e-pub ahead of print], JAMASurg. http://dx.doi.org/10.1001/jamasurg.2017.0904.2017, accessed July 12, 2017.
Assessment for infection risk factors
At pre-operative evaluation, risk factors for infection should be identified (see Section I), and interventions to reduce risk should be implemented when possible. Poor nutritional status is associated with increased risk of MCS-related infection. Nutritional assessment through history and laboratory data (e.g., BMI, serum pre-albumin) is indicated. Consultation with a nutrition expert may be needed.
Other conditions, such as psoriasis, hidradenitis suppurativa, previous sternal radiation therapy, long-term use of immunosuppression or steroids, renal failure and presence of percutaneous endoscopic gastrostomy tube or colostomy, may impact wound healing and should be considered in the care plan. Last, evidence of poor dentition with septic dental foci, such as periodontitis and abscesses, should be addressed before MCS implantation, similar to candidates for organ transplantation, but these procedures are not without morbidity.
81- Rustemeyer J.
- Bremerich A.
Necessity of surgical dental foci treatment prior to organ transplantation and heart valve replacement.
, 82- Smith M.M.
- Barbara D.W.
- Mauermann W.J.
- et al.
Morbidity and mortality associated with dental extraction before cardiac operation.
Therefore, dental evaluation should be included and, if necessary, a dental consultation should be obtained.
83- Feldman D.
- Pamboukian S.V.
- Teuteberg J.J.
- et al.
The 2013 International Society for Heart and Lung Transplantation guidelines for mechanical circulatory support: executive summary.
Allergy
We recommend clarifying the accuracy and significance of all reported anti-microbial allergies. Consultation with an allergy specialist may be useful to confirm or exclude reported antibiotic allergies.
Recommendations:- 1.
Femoral line placement should be avoided if possible.
- 2.
Tunneled catheters are preferred for prolonged use (e.g., dialysis catheters).
- 3.
All vascular access sites (including intra-aortic balloon pump sheaths, arterial lines, Swan--Ganz catheters and PICCs) should be carefully evaluated on a daily basis.
- 4.
Lines should be examined for signs of infection and exchanged if necessary as part of a pre-VAD decontamination strategy.
- 5.
Catheters with evidence of infection should be removed and replaced using a strict sterile technique.
- 6.
Exchange over a wire is strongly discouraged due to risks for bacterial seeding unless done in special circumstances (e.g., no other access).
- 7.
Indwelling urinary catheter placement should be done only if indicated and removed as soon as possible.
- 8.
MRSA screening should be considered, and mupirocin and chlorhexidine used in those patients colonized.
- 9.
Screening for other resistant bacteria should be based on local epidemiologic patterns.
- 10.
Before surgery, the patient should shower or bathe with soap or antiseptic agent at least the night before operation.
- 11.
Dental evaluation before MCS implantation is recommended. If poor dentition with septic foci, such as periodontitis and abscesses, are found, then these should be dealt with before MCS implant.
- 12.
Review antibiotic allergies and obtain specialist assessment as necessary.
Surgical management
The surgical management during MCS implantation may have a major impact on peri- and post-operative infectious complications. Although the surgical implantation procedure of MCS devices is not standardized, and the surgical approach varies based on center, surgeon and device-specific preferences, the basic principles of surgical sterility and infection prevention are universally followed. Nevertheless, there are many common features of the MCS implant procedure and only few that are unique to the specific type of device.
Most MCS devices implanted today are continuous-flow LVAD pumps, which are preferred over the older generation of bulky, pulsatile pumps. Anti-microbial prophylaxis is the standard of care for patients undergoing any cardiac surgical procedure. General recommendations for prevention of SSIs in MCS are presented in the following section, “Anti-microbial prophylaxis”. In general, no difference in surgical sterility is maintained during implantation of the device in comparison to any other cardio-surgical procedure. Although meticulous attention to sterility measures must always be undertaken, extreme measures like laminar flow and helmet usage, as seen in orthopedic surgery, are not routinely practiced in cardiac surgery.
The surgical implant procedure starts with clipping of the chest hair, immediately preceding the operation, followed by appropriate prepping and draping of the patient. Alcohol-based agents should be used for skin preparation in the operating room unless contraindicated.
80Berrios-Torres SI, Umscheid CA, Bratzler DW, et al. Centers for Disease Control and Prevention Guideline for the Prevention of Surgical Site Infection [e-pub ahead of print], JAMASurg. http://dx.doi.org/10.1001/jamasurg.2017.0904.2017, accessed July 12, 2017.
Patients’ potential allergies must be taken into consideration when local disinfectants are used. At many centers, an additional sterile anti-microbial incision drape containing bound iodine is used for coverage of the incision sites. Topical anti-microbial irrigation of the subcutaneous tissue is done after placement of the sternal wires before closure of the sternal wound. This approach is based on data showing that topical use of anti-microbials, mainly gentamicin and vancomycin, in combination with systemic antibiotics may significantly reduce SSIs.
84Modified prophylaxis for preventing deep sternal wound infection after cardiac surgery.
However, 2 more recent randomized studies could not confirm the usefulness of topical antibiotics in cardiac surgery,
85- Eklund A.M.
- Valtonen M.
- Werkkala K.A.
Prophylaxis of sternal wound infections with gentamicin-collagen implant: randomized controlled study in cardiac surgery.
, 86- Bennett-Guerrero E.
- Ferguson Jr, T.B.
- Lin M.
- et al.
Effect of an implantable gentamicin-collagen sponge on sternal wound infections following cardiac surgery: a randomized trial.
and the 2013 Guidelines for Antimicrobial Prophylaxis in Surgery does not recommend their routine use for cardiac procedures.
87- Bratzler D.W.
- Dellinger E.P.
- Olsen K.M.
- et al.
Clinical practice guidelines for antimicrobial prophylaxis in surgery.
The recent CDC guidelines for SSI prevention do not recommend application of anti-microbials in the form of ointment, solution or powder to surgical incisions.
80Berrios-Torres SI, Umscheid CA, Bratzler DW, et al. Centers for Disease Control and Prevention Guideline for the Prevention of Surgical Site Infection [e-pub ahead of print], JAMASurg. http://dx.doi.org/10.1001/jamasurg.2017.0904.2017, accessed July 12, 2017.
During surgery, glycemic control, with blood glucose level target <200 mg/dl, and normothermia are recommended to prevent SSI.
80Berrios-Torres SI, Umscheid CA, Bratzler DW, et al. Centers for Disease Control and Prevention Guideline for the Prevention of Surgical Site Infection [e-pub ahead of print], JAMASurg. http://dx.doi.org/10.1001/jamasurg.2017.0904.2017, accessed July 12, 2017.
An increased fraction of inspired oxygen should be given during surgery and after extubation in the immediate post-operative period, and transfusion of blood products should not be withheld in attempts to prevent SSI.
80Berrios-Torres SI, Umscheid CA, Bratzler DW, et al. Centers for Disease Control and Prevention Guideline for the Prevention of Surgical Site Infection [e-pub ahead of print], JAMASurg. http://dx.doi.org/10.1001/jamasurg.2017.0904.2017, accessed July 12, 2017.
Essential to the surgical approach is to minimize surgical stress and trauma to reduce inflammatory responses related to the procedure. The size of the incision, as well as the duration of surgery, may have a major impact on development of infectious complications after implantation. Many of the devices are directly attached to the heart and/or the aorta, and an open access to these internal structures must be made. Approximately 95% of all VAD implants are performed via median sternotomy, which provides good access to the heart and the ascending aorta. Unfortunately, this approach has been associated with a greater risk of SSI. Due to the shape and size of certain pumps in use today (e.g., HeartMate II), an additional pre-peritoneal pump pocket must be achieved. Other smaller pumps can be placed inside the pericardium, which prevents potential bleeding into the pump pocket and may decrease infection risk related to blood retention and clot formation.
88Risk factors for surgical infection.
In recent years, a less invasive surgical approach for implantation of new-generation pumps (HeartWare HVAD and Jarvik 2000) has been developed. This approach includes a small left-sided mini-thoracotomy in combination with an upper mini-sternotomy or right-sided mini-thoracotomy in place of a full median sternotomy.
89- Anyanwu A.C.
- Itagaki S.
- Pinney S.
- et al.
Initial experience with routine less invasive implantation of HeartMate II left ventricular assist device without median sternotomy.
, 90- Bejko J.
- Guariento A.
- Bortolussi G.
- et al.
Minimally invasive surgical Jarvik 2000 off-pump implantation.
, 91- Haberl T.
- Riebandt J.
- Mahr S.
- et al.
Viennese approach to minimize the invasiveness of ventricular assist device implantation.
However, due to small patient numbers, evidence is still lacking that this limited incision is associated with fewer SSIs compared with the full sternotomy procedure.
Before implantation, the pump is tested in a bath of sterile saline or dextrose, which is performed on a sterile back table. This pump testing is typically performed by a VAD coordinator or perfusionist, simultaneously to the opening of the chest and preparation of the surgical implant site. After testing, the VAD is kept wrapped in a sterile towel until implantation. There is no consensus as to whether the towel should be soaked with an antibiotic solution to prevent contamination during the resting period on the back table before implantation. Also, there is no consensus as to whether the outflow graft and the Dacron-velour coating of the drive-line should also be soaked with antibiotic solution before implantation. This technique of soaking vascular grafts before implantation with antibiotic solutions (mainly rifampin) is practiced by vascular surgeons in some centers throughout the world.
92- Vicaretti M.
- Hawthorne W.J.
- Ao P.Y.
- et al.
An increased concentration of rifampicin bonded to gelatin-sealed Dacron reduces the incidence of subsequent graft infections following a staphylococcal challenge.
The approach has some potential risks, including allergy, toxicity and drug--drug interactions from absorption of these antibiotics into the systemic circulation, and may need careful evaluation. There are no randomized, controlled trials to evaluate the practice of soaking prosthetic devices with antibiotics, and therefore this practice is not currently recommended according to the recent CDC guidelines.
80Berrios-Torres SI, Umscheid CA, Bratzler DW, et al. Centers for Disease Control and Prevention Guideline for the Prevention of Surgical Site Infection [e-pub ahead of print], JAMASurg. http://dx.doi.org/10.1001/jamasurg.2017.0904.2017, accessed July 12, 2017.
To stabilize the subcutaneous course of the drive-line and facilitate tissue ingrowth, all pump drive-lines are partially coated with Dacron-velour. Using a modified surgical tunneling technique, the entire drive-line velour portion is kept in the subcutaneous tunnel, resulting in a silicone--skin interface at the exit site.
93- Dean D.
- Kallel F.
- Ewald G.A.
- et al.
Reduction in driveline infection rates: results from the HeartMate II Multicenter Driveline Silicone Skin Interface (SSI) Registry.
, 94- Singh A.
- Russo M.J.
- Valeroso T.B.
- et al.
Modified HeartMate II driveline externalization technique significantly decreases incidence of infection and improves long-term survival.
This modified surgical approach has shown better long-term post-operative hygiene and easier care of the exit site and the drive-line. It proved to be associated with fewer SSIs than the conventional technique, which leaves a portion of the velour outside the skin, and thus has been adopted by the majority of centers.
3- Kusne S.
- Danziger-Isakov L.
- Mooney M.
- et al.
Infection control and prevention practices for mechanical circulatory support: an international survey.
, 93- Dean D.
- Kallel F.
- Ewald G.A.
- et al.
Reduction in driveline infection rates: results from the HeartMate II Multicenter Driveline Silicone Skin Interface (SSI) Registry.
, 94- Singh A.
- Russo M.J.
- Valeroso T.B.
- et al.
Modified HeartMate II driveline externalization technique significantly decreases incidence of infection and improves long-term survival.
An important step during VAD implantation is determination of the location of the VAD drive-line exit site. Historically, the drive-line exit site was positioned in the right upper abdominal quadrant with a long subcutaneous tunneled course, and many pump manufacturers adapted their length of the drive-line based on this approach. To prevent subcutaneous extension of infection from the drive-line exit site toward the pump or pump pocket, several centers started to increase the subcutaneous tunneled course by creating a “C-shaped” course of the drive-line, exiting on the left lower abdominal quadrant.
95- Schibilsky D.
- Benk C.
- Haller C.
- et al.
Double tunnel technique for the LVAD driveline: improved management regarding driveline infections.
A recent series described the different surgical approaches in continuous-flow VAD implantation including the classical approach with median sternotomy, minimally-invasive without sternotomy, and the lateral implantation to the descending aorta.
96- Krabatsch T.
- Drews T.
- Potapov E.
- Weng Y.
- Pasic M.
- Hetzer R.
Different surgical strategies for implantation of continuous-flow VADs-experience from Deutsches Herzzentrum Berlin.
To reduce the mechanical irritation of the drive-line exit site and promote the healing process in the early peri- and post-operative periods, some centers have adapted the technique of fixation of the drive-line to the skin with 1 or 2 sutures for a period of 2 or 3 weeks. This facilitates healing of the exit site with less mechanical irritation and may prevent early and late infections.
97- Yarboro L.T.
- Bergin J.D.
- Kennedy J.L.
- et al.
Technique for minimizing and treating driveline infections.
Recommendations:- 1.
General principles of infection control and SSI prevention should be followed.
- 2.
Allergy should be taken into consideration when local skin antiseptics are used.
- 3.
Clipping of chest hair is done immediately before surgery.
- 4.
Before implantation, the device is to be tested on the back table under sterile conditions.
- 5.
Skin preparation in operating room should be performed with an alcohol-based agent unless contraindicated.
- 6.
During surgery, glycemic control, with blood glucose target <200 mg/dl, and normothermia should be maintained.
- 7.
Increased fraction of inspired oxygen is important to maintain during surgery and after extubation in the immediate post-operative period.
- 8.
Transfusion of blood products should not be withheld but given if necessary.
- 9.
The velour coverage of the drive-line is kept under the skin at the exit site for better care of the drive-line exit site.
- 10.
To lessen shearing and foster granulation along the drive-line exit site, the drive-line should be secured in the immediate post-operative period.